CVE-2015-3954: Hospira Plum A+ Infusion System vulnerability
Hospira · Plum A+ Infusion System
Affected: Hospira Plum A+ Infusion System · Exploitation: possible (1 to 10%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 81% of scored CVEs
Not scored yet. The CVE record carries no CVSS score.
DescriptionCVE.org
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Affected products3
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- Hospira · Plum A+ Infusion System<= 13.4
- Hospira · Plum A+3 Infusion System<= 13.6
- Hospira · Symbiq Infusion System<= 3.13
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
- Exploitation
- Possible (1 to 10%). EPSS between 1% and 10%.
- CISA KEV
- Not listed
- Public check
- No public nuclei template
- Vendor
- Hospira
- Published
- March 2019
CoverageNo articles yet