CVE-2017-9658: Philips IntelliVue MX40 Patient Worn Monitor vulnerability

Philips · IntelliVue MX40 Patient Worn Monitor

Affected: Philips IntelliVue MX40 Patient Worn Monitor · Exploitation: unlikely (under 1%)

Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.

0.8%probability of exploitation in the next 30 days
57th percentile
050100

Higher than 57% of scored CVEs

Not scored yet. The CVE record carries no CVSS score.

DescriptionCVE.org

Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Affected products1

From the CVE record. A product or version that is not listed is unknown, not unaffected.

  • Philips · IntelliVue MX40 Patient Worn MonitorIntelliVue MX40 Patient Worn Monitor (WLAN only), all versi…

On the Atlas

Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.

Open the Atlas
Exploitation
Unlikely (under 1%). EPSS under 1%, or no score yet.
CISA KEV
Not listed
Public check
No public nuclei template
Vendor
Philips
Published
April 2018

CoverageNo articles yet

No CyberSIXT article names CVE-2017-9658 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.

References

Updated AboutDevelopersSign in