CVE-2018-10601: ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors vulnerability
ICS-CERT · IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors
Affected: ICS-CERT IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors · Exploitation: unlikely (under 1%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 31% of scored CVEs
Not scored yet. The CVE record carries no CVSS score.
DescriptionCVE.org
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.
Affected products1
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- ICS-CERT · IntelliVue Patient Monitors, Avalon Fetal/Maternal MonitorsThe following IntelliVue Patient Monitors versions are affe…
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
CoverageNo articles yet
No CyberSIXT article names CVE-2018-10601 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.