CVE-2018-8868: Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Medtronic · 24950 MyCareLink Monitor

Vendor fix available · Affected: Medtronic 24950 MyCareLink Monitor · Exploitation: unlikely (under 1%)

Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.

0.3%probability of exploitation in the next 30 days
27th percentile
050100

Higher than 27% of scored CVEs

6.2MediumCVSS 3.1
LowMedium

DescriptionCVE.org

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

Affected products2

From the CVE record. A product or version that is not listed is unknown, not unaffected.

  • Medtronic · 24950 MyCareLink MonitorAll versions
  • Medtronic · 24952 MyCareLink MonitorAll versions

On the Atlas

Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.

Open the Atlas
Exploitation
Unlikely (under 1%). EPSS under 1%, or no score yet.
CISA KEV
Not listed
Public check
No public nuclei template
Vendor
Medtronic
Published
July 2018

CoverageNo articles yet

No CyberSIXT article names CVE-2018-8868 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.

References

Updated AboutDevelopersSign in