CVE-2019-13539: Medtronic Valleylab FT10 and FX8 Reversible One-way Hash
Medtronic · Valleylab Exchange Client
Affected: Medtronic Valleylab Exchange Client · Exploitation: unlikely (under 1%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 19% of scored CVEs
DescriptionCVE.org
Medtronic Valleylab Exchange Client version 3.4 and below, Valleylab FT10 Energy Platform (VLFT10GEN) software version 4.0.0 and below, and Valleylab FX8 Energy Platform (VLFX8GEN) software version 1.1.0 and below use the descrypt algorithm for OS password hashing. While interactive, network-based logons are disabled, and attackers can use the other vulnerabilities within this report to obtain local shell access and access these hashes.
Affected products3
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- Medtronic · Valleylab Exchange Clientup to and including 3.4
- Medtronic · Valleylab FT10 Energy Platform (VLFT10GEN)up to and including software version 4.0.0
- Medtronic · Valleylab FX8 Energy Platform (VLFX8GEN)up to and including software version 1.1.0
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
- Exploitation
- Unlikely (under 1%). EPSS under 1%, or no score yet.
- CISA KEV
- Not listed
- Public check
- No public nuclei template
- Weakness
- Use of Weak Hash · Cryptography
- Vendor
- Medtronic
- Published
- November 2019
CoverageNo articles yet
No CyberSIXT article names CVE-2019-13539 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.