CVE-2019-25719: Dräger Infinity M540 VG4.1.1 Spoofing and DoS via Network Message Handling
Dräger · Infinity Acute Care System
Vendor fix available · Affected: Dräger Infinity Acute Care System · Exploitation: unlikely (under 1%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 2% of scored CVEs
DescriptionCVE.org
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Affected products2
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- Dräger · Infinity Acute Care Systemup to and including VG4.1.1; up to and including VG4.0.3
- Dräger · Standalone Infinity M540 patient monitorup to and including VG4.1.1
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
CoverageNo articles yet
No CyberSIXT article names CVE-2019-25719 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.