CVE-2024-48966: Life2000 service tools for test and calibration do not support user authentication
Baxter · Life2000 Ventilation System
Affected: Baxter Life2000 Ventilation System · Exploitation: unlikely (under 1%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 53% of scored CVEs
DescriptionCVE.org
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Affected products1
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- Baxter · Life2000 Ventilation System06.08.00.00 and prior
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
- Exploitation
- Unlikely (under 1%). EPSS under 1%, or no score yet.
- CISA KEV
- Not listed
- Public check
- No public nuclei template
- Vendor
- Baxter
- Published
- November 2024
CoverageNo articles yet
No CyberSIXT article names CVE-2024-48966 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.