CVE-2025-0683: Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
Contec Health · CMS8000 Patient Monitor
Affected: Contec Health CMS8000 Patient Monitor · Exploitation: unlikely (under 1%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 56% of scored CVEs
DescriptionCVE.org
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
Affected products1
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- Contec Health · CMS8000 Patient MonitorAll versions
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
- Exploitation
- Unlikely (under 1%). EPSS under 1%, or no score yet.
- CISA KEV
- Not listed
- Public check
- No public nuclei template
- Vendor
- Contec Health
- Published
- January 2025
CoverageNo articles yet
No CyberSIXT article names CVE-2025-0683 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.
References
- NVD entrynvd.nist.gov
- CVE recordcve.org
- Weakness CWE-359cwe.mitre.org
- cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01cisa.gov
- fda.gov/medical-devices/safety-communications/cybersecurity-vulnfda.gov
- cisa.gov/resources-tools/resources/contec-cms8000-contains-backdcisa.gov
- bleepingcomputer.com/news/security/backdoor-found-in-two-healthcbleepingcomputer.com