CVE-2018-14781: Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
Medtronic · MMT- 508 - MiniMed pump
Affected: Medtronic MMT- 508 - MiniMed pump · Exploitation: unlikely (under 1%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 54% of scored CVEs
DescriptionCVE.org
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
Affected products10
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- Medtronic · MMT- 508 - MiniMed pumpAll versions
- Medtronic · MMT – 511 pump ParadigmAll versions
- Medtronic · MMT – 512 / MMT – 712 Paradigm x12All versions
- Medtronic · MMT – 515 / MMT – 715 Paradigm x15All versions
- Medtronic · MMT – 522 / MMT – 722 Paradigm REAL-TIMEAll versions
- Medtronic · MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIMEAll versions
- Medtronic · MMT – 523 / MMT – 723 Paradigm RevelAll versions
- Medtronic · MMT – 523(K) / MMT – 723(K) ParadigmAll versions
- Medtronic · MMT – 554 / MMT – 754 MiniMed VeoAll versions
- Medtronic · MMT – 551 / MMT – 751 MiniMed 530GAll versions
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
- Exploitation
- Unlikely (under 1%). EPSS under 1%, or no score yet.
- CISA KEV
- Not listed
- Public check
- No public nuclei template
- Vendor
- Medtronic
- Published
- August 2018
CoverageNo articles yet
No CyberSIXT article names CVE-2018-14781 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.