CVE-2020-25183: Medtronic MyCareLink Smart Improper Authentication
Medtronic · Smart Model 25000 Patient Reader
Vendor fix available · Affected: Medtronic Smart Model 25000 Patient Reader · Exploitation: unlikely (under 1%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 57% of scored CVEs
DescriptionCVE.org
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
Affected products1
From the CVE record. A product or version that is not listed is unknown, not unaffected.
- Medtronic · Smart Model 25000 Patient ReaderAll versions
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
- Exploitation
- Unlikely (under 1%). EPSS under 1%, or no score yet.
- CISA KEV
- Not listed
- Public check
- No public nuclei template
- Vendor
- Medtronic
- Published
- December 2020
CoverageNo articles yet