Skip to content
Sign in
CVE-2020-251838.0HighCWE-287

CVE-2020-25183: Medtronic MyCareLink Smart Improper Authentication

Medtronic · Smart Model 25000 Patient Reader

Vendor fix available · Affected: Medtronic Smart Model 25000 Patient Reader · Exploitation: unlikely (under 1%)

Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.

0.8%probability of exploitation in the next 30 days
57th percentile
050100

Higher than 57% of scored CVEs

8.0HighCVSS 3.1
LowHigh

DescriptionCVE.org

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

Affected products1

From the CVE record. A product or version that is not listed is unknown, not unaffected.

  • Medtronic · Smart Model 25000 Patient ReaderAll versions

On the Atlas

Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.

Open the Atlas
Exploitation
Unlikely (under 1%). EPSS under 1%, or no score yet.
CISA KEV
Not listed
Public check
No public nuclei template
Vendor
Medtronic
Published
December 2020

CoverageNo articles yet

No CyberSIXT article names CVE-2020-25183 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.

References

Updated AboutDevelopersSign in