Skip to content
Connect AISign inSign up free
CVE-2021-37162Not scoredCWE-120

CVE-2021-37162: Swisslog-Healthcare Hmi-3 Control Panel Firmware vulnerability

Swisslog-Healthcare · Hmi-3 Control Panel Firmware

Affected: Swisslog-Healthcare Hmi-3 Control Panel Firmware · Exploitation: possible (1 to 10%)

Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.

3.3%probability of exploitation in the next 30 days
88th percentile
050100

Higher than 88% of scored CVEs

Not scored yet. The CVE record carries no CVSS score.

DescriptionCVE.org

A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.

On the Atlas

Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.

Open the Atlas
Exploitation
Possible (1 to 10%). EPSS between 1% and 10%.
CISA KEV
Not listed
Public check
No public nuclei template
Published
August 2021

CoverageNo articles yet

No CyberSIXT article names CVE-2021-37162 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.