CVE-2021-37166: Swisslog-Healthcare Hmi-3 Control Panel Firmware vulnerability
Swisslog-Healthcare · Hmi-3 Control Panel Firmware
Affected: Swisslog-Healthcare Hmi-3 Control Panel Firmware · Exploitation: possible (1 to 10%)
Not in CyberSIXT coverage yet: no article names this CVE. This page shows the public record.
Higher than 81% of scored CVEs
Not scored yet. The CVE record carries no CVSS score.
DescriptionCVE.org
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.
On the Atlas
Every published CVE is on the CyberSIXT Atlas, grouped by weakness, vendor and month.
- Exploitation
- Possible (1 to 10%). EPSS between 1% and 10%.
- CISA KEV
- Not listed
- Public check
- No public nuclei template
- Vendor
- Swisslog-Healthcare
- Published
- August 2021
CoverageNo articles yet
No CyberSIXT article names CVE-2021-37166 yet, so there is no coverage timeline, action brief or change log for it. Everything on this page comes from the public CVE record, CISA KEV, EPSS and the nuclei template index. Search the archive or see the CVEs in the news.
References
- NVD entrynvd.nist.gov
- CVE recordcve.org
- Weakness CWE-120cwe.mitre.org
- swisslog-healthcare.comswisslog-healthcare.com
- armis.com/PwnedPiperarmis.com
- swisslog-healthcare.com/en-us/customer-care/security-informationswisslog-healthcare.com
- swisslog-healthcare.com/-/media/swisslog-healthcare/documents/cuswisslog-healthcare.com